The device KMP DFM 800 is the easy portable device developed for performance of expanded functions of monitoring and procedures during bad heart attack with use of 4 modules for work: Monitor, Manual defibrillator, automatic defibrillator and pacemaker (option).
Device and principle of work:
- The supervision mode – is recommended for supervision of the vital parameters, including 12 assignments receiving and transfers for the equipment having a similar option.
- The mode of a manual defibrillation – is recommended for a manual defibrillation, synchronized cardioversion and supervision of the vital signs.
- The mode of an automatic defibrillation – is recommended for the analysis of a signal of an ECG and a defibrillation in case of need, with instructions on use for the user.
- The noninvasive mode of cardiostimulation – for the devices offering full function, this mode is recommended for cardiostimulation of pulse and supervision of the vital signals.
- The configuration mode – can be applied to performance of manual checks, a configuration of the different modes, access to information saved on a memory card and performance of functions on service.
- The demo the mode – is used for demonstration of work of the equipment with display of simulation of curves. This mode is connected only in the configuration mode in service options.
In the mode of supervision of DFM 800 can show up to 3 curve electrocardiograms (only for the equipment with this function), to carry out a signal through 4, 5 or 10 assignments of a cable of the patient or by means of external one-time plates for adults or children or through one-time multifunction electrodes. In addition in this mode it is possible to observe oksimetriya module parameters (SpO2, SpMet, SpCO, PVI, SpHb or SpOC), the noninvasive arterial pressure (NIAD) and parameters of the mode of a kapnografiya (etCO2 and AwRR) on the screen, to change parameters of limits for different observed parameters, to receive and transmit signals of 12 assignments of an ECG and to show the report on work and schedules of observed vital parameters.
In the mode of a manual defibrillation if the Category is necessary for the patient, then it is easy to make it three operations:
- Choice of level of energy
- Charging
- Category
In the mode of a manual defibrillation the defibrillator can be carried out by means of reusable external plates, multifunction one-time electrodes or internal plates. It is also possible to carry out a cardioversion synchronization. Access to this operating mode can be limited in configuration options.
During the work in the semi-automatic mode the device analyzes an ECG of the patient and defines whether it is necessary to defibrillate a cordial rhythm and if it is necessary, then will provide to the user of team for work of the Category. During all process the device shows messages on the display and transmits voice messages through the special device the speaker located on the front panel of the device. In the mode of an automatic defibrillator the device can be configured for record of audio of the signal given during connection and work.
The pacemaker mode (additional) provides noninvasive endocardial stimulation delivering a rhythm through one-time electrodes of multifunction devices.
The device KMP DFM 800 can work at NiMH batteries which can be recharged, also it can be connected to the alternating current main or to the automobile accumulator. The indicator of the status of the battery constantly is shown in upper part of the display. Also at connection of the device to an external energy source the battery is automatically recharged irrespective of, the device is connected or not.
The device has a recorder of high resolution which the user can configure, it is possible to unpack forms of waves and operating modes. The user can change settings of a recorder.
In addition to these modes oriented to the patient the device has the special starting mode which provides direct access to the configuration mode in which the user can change and configure settings of control of operations of the device for convenience of work with the device. The device can also carry out various tests at the request of the user, using configuration mode options.
The device KMP DFM 800 carries out a series of tests on self-examination at turning on of the device and in operating time that allows to find malfunctions in system of work or some problems which can lead to interruptions in operation of the device and will make it unsafe. The indicator of malfunctions located on the front panel of the device shows the found mistake, also messages on it appear on the display.
Also the device KMP DFM 800 has an option of automatic saving information on an external memory card (flash carrier). These data include an ECG of the patient, an event, the actions happening in a usage time of the device and audiosignals (in addition) and also background noise if the device works in the mode of an automatic defibrillation. In addition to this information 100 last actions which happened in a device usage time are stored in memory, they are grouped in use options.
Technical characteristics:
MONITOR | |
ECG | Supervision by means of 4, 5 and 10 assignment of a cable, reusable internal and external plates and one-time multifunction electrodes |
Assignments | 4 assignment cable: PLATES, I, II, III, aVR, aVL and aVF 5 assignment cable: PLATES, I, II, III aVR, aVL, aVF and V 10 assignment cable: PLATES, I, II, III, aVR, aVL, aVF and V1 to V6 |
The indicator without assignments | On the display the badge at shutdown or bad connection of assignment will appear Amplitude at present connected to the patient for definition without assignment less than 0.5 uA. |
ECG size | 0.5, 1, 2 and 4 cm/m V chosen with per-it panels |
ECG speed on the display | 25 mm/sec. |
Frequency response | AC filter (50/60 Hz). diagnostics: 0.05-150 Hz (only on the printer) Filter muscular sokrashcheniyr: 0.67-40 Hz (only on the printer) Display response: 0.05-25 Hz |
ChSS (heart rate) | 30-300 bpm ± 10% for show on the display of the device |
Average ChSS | For ChSS equal or it is more than 50 bpm, 8 latest R-R of intervals are applied to alignment of ChSS For ChSS it is less than 50 bpm, 4 latest R-R of an interval priimenyatsya for alignment of ChSS |
ChSS response time | From 80 to 40 bpm: 3 seconds from 80 to 120 bpm: 2 seconds |
Alarm signal response time for tachycardia |
|
Ability to reject T-waves | Rejects T-waves with the maximum amplitude of 0.7 mV |
Alarm signals |
|
Suppression of a cophased component | > 100 dBs |
SpO2 pulsoksimetriya (option) | |
Fullness (%SpO2) accuracy level | 1% - 100% |
Fullness (SpO2%) in the conditions of an immovability | the adult / pediatrich 70% - 100%: ± 2 figures 0% - 69%: it is not specified Newborn 70% - 100%: ± 3 figures 0% - 69%: it is not specified |
Fullness (%SpO2) accuracy on condition of the movement | the adult / pediatrich / 70% - 100%: ± 3 figures novorozhd. 0% - 69%: it is not specified |
Fullness (SpO2%) permission | 1% |
Range of pulse rate (bpm) | 25-240 bpm |
Pulse rate (ppm) accuracy on condition of lack of the movement | ~ 2 bpm |
Pulse rate (ppm) accuracy on condition of the movement | ~ 5 bpm |
Pulse rate (ppm) permission | 1 bpm |
DEFIBRILLATOR | |
Signal phase | Two-phase truncated exponential, with compensation of energy according to resistance of the patient of impedance |
Energy output accuracy (more than 50 Ω) | ~ 15% or ~ 3 J that is the greatest for all range |
Manual defibrillator | |
Energy at the exit |
1 - 2 - 3 - 5 - 7 - 9 - 10 - 15 - 20 - 30 - 50 - 70 – 100-125 - 150 - 200 Joules 1 - 2 - 3 - 5 - 7 - 9 - 10 - 15 - 20 - 30 - 50 Joules |
Options of plates | Reusable external plates Internal plates Multifunction one-time electrodes Constant cable for one-time multifunction electrodes |
Energy choice | The button on the front panel and buttons on external plates |
Loading control | The button on the front panel and buttons on external plates |
Loading indicator | The prizaryazhaniya sound, sound at the end of loading, ZhKD charging button, the button the category on the front panel blinks for one-time multifunction electrodes and internal plates |
Control of the category | Buttons on external plates, the button on the front panel for one-time multifunction electrodes and internal plates |
Charging time | Less than 5 sec. at 200 J with the new and completely charged NiMH battery at 25 0C. Less than 10 sec. without battery, it is connected to energy to a voltage at 90-100% of nominal rate Less than 10 sec. with the new and completely charged with NiMH battery, with the carried-out 15 categories on 200 J at 25 0C. |
The maximum time from the beginning of connection of an energy source to readiness for the category | Less than 10 sec. from the moment of inclusion with the new and completely charged NiMH battery. Less than 15 sec. from the moment of inclusion, without battery, it is connected to the power supply for 90-100% of nominal rate Less than 15 sec. with the new and completely charged NiMH battery with the carried-out 15 categories on 200 J at at 25 0C. |
Synchronization | The button on the front panel. Display on the display of points of synchronization |
The maximum delay time between an impulse of synchronization and delivery of energy | Each delivery of energy happens during 60 ms right after definition of peak of QRS |
Semi-automatic defibrillation (option) | |
Energy of an exit | At most: 200 J ~ 15% |
Plate option | One-time multifunction electrodes the Constant cable with one-time electrodes |
Messages of the recommendation | Removal on the screen and audio of the message which help the user in operating time |
Charge indicator | Charging sound, sound Concha of charging and a button blinking the category on the front panel |
Control of the category | Button of the front panel |
Configuration of the applied parameters | By means of the corresponding options of the mode of a konf-tion |
Properties of definition | VF sensitivity: compliance to standards of safety of AHA VT sensitivity: compliance to standards of safety of AHA NSR specifics: compliance to standards of safety of AHA Specifics of other signals: compliance to standards of safety of AHA |
The maximum time from the beginning of the analysis of a cordial rhythm until readiness for the category | Less than 20 sec. with the new and completely charged NiMH batteries Less than 20 sec. without battery, it is connected to energy at 90-100% of nominal rate Less than 20 sec. with new and a polnost the charged NiMH batteries with the carried-out 15 categories on 200 J at 25єC. |
The maximum time from initial energy supply to readiness for the category | Less than 26 sec. with the new and completely charged NiMH batteries Less than 26 sec. without battery, it is connected to energy at 90-100% of nominal rate Less than 26 sec. with new and a polnost the charged NiMH batteries with the carried-out 15 categories on 200 J at 25єC |
PACEMAKER (Option) | |
Signal phase | Continuous priyamolineyny form |
Pulse width | 40 ms |
Amplitude | From 0 to 150 mA with a step to 5 mA |
Speed | from 30 to 180 bpm with a step to 5 bpm |
Operating modes | fixed as necessary |
Dispersion/refraction period | 240 ms from 30 to 80 bpm 340 ms from 85 to 180 bpm |
Automatic press | It is possible to make a configuration of the automatic press for 8 seconds before and after events which disconnect preventions and a defibrillation |
Reports | Use of the report on work; schedules of ChSS and graphics of SpO2 of % (option); Redultata of checks by the user and self-examinatione of the device; Configuration Parameters; The Events kept on a memory card together with an ECG corresponding to signals |
Paper width | 50 mm |
Speed | 10,25 and 50 mm/sec. ± 5% |
DATA STORAGE (OPTION) | |
Memory type | The external taken-out USB stick/memory card |
Memory size | At least 16 MB that is equal to 4 hours of continuous record of signals of an ECG plus audio |
Data | Constant signal of an ECG plus audio (option) Important events together with a sottvetstvuyushchy signal of an ECG |
THE GENERAL | |
Indicators | Indicator of the status of the battery Indicator of malfunctions of the device Energy supply indicator Charge indicator Indikatornakopleniya of energy Synchronization indicator |
Self-examination of the device | At inclusion In operating time Separately at the request of the user |
ENERGY SUPPLY | |
Battery Type Power
|
NiMH (loaded) More than 130 categories in 200 J at 20єC More than 150 minutes of monitoring, More than 120 minutes of monitoring plus pacemaker (60 mA and 60 bpm) About 3 hours 800 gr |
Alternating current | 100-240 V AC and 50-60Hz |
Constant (car accumulator) | 10-16 V DC |
Equipotential conductor | Provides additional connection to grounding of the building where the device is installed. If this grounding is unavailable, then connect the equipotential connector to any metal to an element in the building |
ENVIRONMENTAL CONDITIONS | |
Teaperatura of work | From 0єC to 50єC only in the mode of monitoring of a defibrillation, with established batteries and without supply of electricity From 0єC to 40єC at connection to an electric power source |
Storage temperature | From 20єC to 60єC except the battery and one-time multifunction electrodes |
Relative humidity | From 10 to 95% |
The atmospheric pressure (existing) | Surrounding up to 525 mmHg (0 to 3,000 m) |
Vodonepronichayemost | IPX2 |
Vibration | IEC 60068-2-64 |
Category | IEC 60068-2-27 |
PHYSICAL CHARACTERISTICS | |
Weight | the device with the printer, reusable external plates and the battery: 6.9 kg The device with the printer, reusable multifunction electrodes and the battery: 6.0 kg The device with a recorder, the option SpO2, AED, pacemaker, reusable multifunction electrodes and battery: 6.3 kg reusable external plates: 0.95 kg battery: 0.8 kg |
Sizes | 195 mm height x 249 mm length x 310 mm width |