Defibrillator - monitor DKI - n - 11 "AXION" used in medical institutions and on departures ambulance brigades. Multi - function device lets you watch and record EKG patient carries out pulsoksimetriju and measures blood pressure. Defibrillator DKI - n - 11 has a colour TFT display 7 "(800? 640 pixels), built - in thermal printer, battery charger and replaced memory card. The total weight of a defibrillator will not exceed 7 kg.
Equipment:
- Full: defibrillator + ECG + SpO2 + NIBP + JeKSN + memory card
- Minimum: defibrillator + ECG + memory card
Technical characteristics of defibrillator DKI - n - 11 "AXION":
Pulse energy:
- for adults 5, 10, 25, 50, 75, 100, 150, 200, 250, 300, 360 Joules (energy set operation over 200 Joules when dealing with older electrodes)
- for children 5, 10, 25, 50, 75, 100, 150 Joules (blocking issuance of energy more than 150 Joules in baby mode)
Set time power
- 200 J - no more than 6 with
- 360 J: not more than 10 with
The number of digits of the energy from the fully charged battery:
- 200 J: not less than 70
- 360 J: not less than 40
Time of continuous operation in monitoring mode:
- replacement of the battery: no less than 3: 00.
- DC power supply (12 - 18) in not less than 168 hours
Channels:
- Channel 3/6/12 - lead ECG EKG
- Channel NIBP - HELL dimension in unattended mode in the range from 20 to 280 mmHg with an absolute error of ± 3 mm Hg.
- Protection of the maximum pressure, more than 330 mmHg
- SpO2 channel - range of SpO2 readings ranging from 0 to 100%
- Range of SpO2 readings ranging from 75 to 100% with a tolerance of ± 3% of the ex - Channel external, jendokardialnyj and chrespishhevodnyj
- Ex - channel work modes: fixed sverhstimuljacija Fixed, Overdrive, and on - demand Broadcasts.
Installation:
- frequency setting: from 40 to 250 pulses/min,
- setting the pulse width: from 20 to 40 MS,
- setting the current pulse: from 10 up to 180 Ma, for external stimulation.
Defibrillation:
- When the outside stimulation used disposable electrodes.
- Impulse is bi - phasic, trapezoidal shape, unbalanced, with negative and positive current poluvoln (0.5 ± 0.1)
- Blocking issuance of energy less than 12 Ohms and more than 200 Ohm
- Automatic current limiting defibrillation on the level (30 ± 10) and when the patient's body less than 25 Ohms
- Automatic stabilisation of output pulse parameters from 25 - 175 Om
- Verbal and Visual tracking of operator activities.
|