Revolutionary technologies in fight against a sudden cardiac death
When using a bi - phase impulse electric current once changes the direction during the category. Thus, current actually passes through heart twice. It allows to reduce almost twice current in comparison with the standard monophase impulse. Besides, the technology of stabilization of the current (CCD), provides invariable current throughout all category that does an impulse almost rectangular and eliminates the peak fluctuations of current inherent to other pulse shapes (exponential, trapezoid, sinusoidal, etc. ). Use of these technologies allows to reach maximum efficiency of a defibrillation and to minimize the damaging impact of electric current on a myocardium.
Distinctive characteristics
- Squared bi - phase defibrillatory impulse with technology of stabilization of the current (CCD)
- Asynchronous / Synchronized (cardioversion) an outside defibrillation in the manual mode
- Time of a set of energy of 200 J in the manual mode
- Children's and adult reusable defibrillatory electrodes - "irons"
- Mode of the Semi - automatic Outside Defibrillation (ANDES) *
- Highly sensitive algorithm of identification of the fibrillation of ventricles (FV) and the ventricular tachycardia (VT) in the mode of the ANDES *
- Hints (voice and text) on carrying out reanimation and a defibrillation in Russian in the mode of the ANDES *
- Chrezkozhny electrocardiostimulation (EX - ) *
- One - time self - adhesive electrodes for a defibrillation / EX - /an ECG
- Big high - contrast monochrome ZhK TFT display
- Registration of 6 assignments of an ECG and display on the display of 2 assignments of an ECG
- Built in pulsoksimetr *
- Built - in thermal printer
- The simple user interface (control of one button in the mode of the ANDES * )
- The Russian - language menu and voice hints on work with the device in Russian
- The CF replaceable memory card for data storage (including an ECG)
- Preservation of an ECG and dictation records of negotiations lasting up to 75 watch on a memory card of 2 GB
- The program for viewing of an ECG and other data from a memory card on the personal computer for the subsequent analysis and documentation
- Built - in block of food/charging
- Power supply from the removable recharged accumulator
- Indicator of a condition of the device and accumulator
- Self - testing function
- Shock - resistant design with internal elements in udaropogloshchayushchy foam material
- Compactness and small weight
- Warranty period 2 years on the device (6 months on the accumulator)
Defi Monitor XD model |
XD1 |
XD3 |
XD10 |
XD30 |
XD100 |
XD110 |
XD300 |
XD330 |
Defibrillation in the manual mode |
▪ |
▪ |
▪ |
▪ |
▪ |
▪ |
▪ |
▪ |
Defibrillation in the mode of the ANDES |
|
|
|
|
▪ |
▪ |
▪ |
▪ |
Pulsoksimetriya (SpO2) |
|
▪ |
|
▪ |
|
|
▪ |
▪ |
EX - (chrezkozhny electrocardiostimulator) |
|
|
▪ |
▪ |
|
▪ |
|
▪ |
Defibrillation: |
|
Pulse shape: |
bi - phase, rectangular, with stabilization of current |
Manual mode: |
- an asynchronous defibrillation / the defibrillation (cardioversion) synchronized with an ECG |
Energy levels |
5/10/20/30/50/100/200/300/360 J |
Set time |
|
Mode of the ANDES: * |
- semi - automatic outside defibrillation * |
Energy levels in the AND * mode: |
a series from three categories with automatic increase of level of energy (140 – 360 J depending on an impedance |
Electrodes: |
- reusable electrodes - "irons" (the adult and built - in them nurseries) |
Chrezkozhny electrocardiostimulator (EX - ): * |
|
Pulse shape: |
bi - phase, rectangular, lasting 20 ms |
Operating modes: |
Fix (EX - with a constant frequency) / Demand (EX - on demand) / Overdrive (EX - ultrahigh frequency for stopping of supraventricular paroxysmal arrhythmias) |
Frequency range: |
Fix/Demand: 30 - 180 / mines, Overdrive: 30 - 250 / |
Stimulation by current: |
30 - 180 ma with linear increase to a preset value during the first 4 impulses, except the Demand mode (soft start technology) |
Monitor: |
|
High - contrast monochrome TFT LCD display |
|
Permission: |
320 x 240 |
Pixel size: |
0.36 x 0.36 mm |
Display size: |
115 x 86 mm |
Diagonal: |
144 mm / 5,7" |
Display: |
2 ECG channels, 1 channel of a curve of pulse * , ChSS, SpO2 * , quantity of defibrillations, quantity of episodes of ZhF and ZhT, reanimation duration, battery charge |
ECG: |
|
Assignments: |
2 of 6 (I, II, III, aVR, aVL, aVF) |
Patient's resistance: |
23 – 200 Ohms |
ChSS range: |
30 – 300 beats/min |
Alarm of ChSS: |
30 – 300 beats/min |
Protection against a defibrillation: |
built - in |
Pulsoksimetriya: * |
|
Type: |
NELLCOR OxiMax SpO2 |
Measuring range: |
0 – 100% Alarm of SpO2 with changeable values |
Printer: |
|
Built - in thermal printer |
|
Channels of the press: |
1 – 6 |
Thermal paper width: |
58 mm |
Broach speed: |
25/50 mm/sec. |
Permission: |
200 dpi |
Press modes: |
AUTO, ONLINE, MEMO |
Food: |
|
The accumulator recharged by NiCd: |
12 V / 1.4 Ach |
Monitoring of an ECG from the accumulator: |
4 hours |
Categories on 200 J from the accumulator: |
70 |
Accumulator charging time: |
2.5 hour |
Built - in block of food/charging: |
alternating current main of 90 - 264 V, 50/60 Hz |
Data processing: |
|
Audio recording, record ECG and other data on a memory card |
|
Memory card of CompactFlash (CF): |
from 512 MB |
Till 75 o'clock record ECG and an audio recording on the card of 2 GB |
|
The program for viewing of data from a memory card on the personal computer |
|
General characteristics: |
|
Dimensions (SHKHVKHG): |
29х33х16 cm |
Weight with the accumulator: |
5.9 kg |
Guarantee of the producer: |
2 years on the device (6 months on the accumulator) |