(View map)
High-performance surgical system like S-arch it is intended for roentgenoscopic visualization and visualization of an aim X-ray analysis of patients in the course of diagnostic, surgical and intervention procedures. The holangiografiya, endoscopy, an urological, orthopedic, neurologic, vascular, cordial, intensive care and procedures of acute management can be examples of a clinical use. The system can be used for other uses of visualization at the discretion of the doctor.
Features:
- Completely automated S-arch (lightweight and simple movement).
- The special manual S-arch of our production provides faster adjustment of height
- Special function "Doctor Handle"
- There is a possibility of equipment of the additional button of the fixing located on the amplifier of the image allowing to establish the S-arch on the party of the doctor (I.I. party). By means of this function there is no need every time to return to the basic provision of the S-arch for movement of the S-arch.
- C-memory function
- This function allows to write down in memory and to reproduce 3C [Rotation of the camera, collimation, roentgenoscopy conditions] for two provisions of the S-arch. By means of this function time and a dosage at radiation of a hip joint is preserved, etc. which is carried out with continuous use of two provisions of the S-arch: front/lateral.
● the S-arch without visible wires
Lack of cables or levers allows to clear the S-arch quickly.
● Memory for 172 pictures
There is a possibility of preservation to 172 pictures such as LIH and for roentgenoscopy which can be seen during operation.
● the Roentgenoscopic picture on the touch panel
This picture can be displayed on the S-arch wheelchair.
● Standard and expanded modes
Are chosen depending on preferences of the operator.
● Folding monitor
Movement up/down, inclination.
● Pulse roentgenoscopy
Standardly well to 15 frames per second. High quality of the image remains at radiation reduction.
● Virtual collimation
During the work with keeping of the last picture virtual collimation allows not to use pictures for collimation and situation.
● Display in real time of a necessary dosage
Monitoring of a dosage in real time at operation.
● Small resistance when using castors
Small castors provide smooth movement of a wheelchair for the S-arch.
● Function of the assistant (User's guide)
Functions of assistant/software configuration of default/calibration of the touch panel are available. If the operator does not know how to use these or those buttons, he can look at the screen with the description of all buttons.
Technical characteristics:
Mobile S-arch | |
SED (Radiator entrance distance) | 980mm |
Min-y the horizontal level of a X-ray from a floor | 775mm; 1105mm (With the S-arch under a bed) |
Rotation of the S-arch | 120 ° |
Axial rotation of the S-arch | 300/-120 ° |
Movement of the S-arch top / down | 450mm |
Movement of the S-arch forward/back | 200mm |
Turn of the S-arch | ±12.5 ° |
Free space of the S-arch | 780mm |
X-ray controller | |
Rated max. power | 2kW |
Range of installation of tension of a tube | 40 to 110kV |
X-ray analysis (film) | |
Short-term rated power | 60kV, 200mAs (20mA); 110kV, 100mAs (12mA) |
MAs installation range | 0.5 to 200mAs |
DR (Digital X-ray analysis) | |
Short-term rated power | 110kV, 16mA |
Pulse roentgenoscopy | |
Speed of pulse roentgenoscopy | 3.75, 7.5, 10, 15fps |
Nome-noye max. tension of a tube and max. current of a tube | 110kV, 5mA (200V cut.); 110kV, 3mA (100V cut.) |
Max. current of a tube and max. tension of a tube for achievement of max. current of a tube | 9mA, 60kV (200V mode); 5mA, 80kV (100V mode) |
Long roentgenoscopy | 30fps |
The strengthened pulse roentgenoscopy | |
Speed of the strengthened pulse roentgenoscopy | 3.75, 7.5 fps |
Rated max. voltage of a tube and max. current of a tube cat. can pass with a rated max. voltage | 110kV, 9.8mA |
Max. current of a tube and max. tension of a tube for achievement of max. current of a tube | 13.5mA, 80kV |
X-ray tube | |
X-ray tube type | Stationary anode |
Focus point size | 0.6mm |
Heat capacity of the anode | 100kHU |
Heat capacity of the mechanism of a tube | 680kHU (480kJ) |
Max. speed of cooling of a tube | 226HU/s (160W) |
Purpose (corner and material) | 8 °, Tungsten |
Collimator | |
Octagonal collimator | Electric drive (otkr./zakr.), it is made of W |
Parallel collimator | Electric drive (otkr/zakr. and vrashch.), it is made of Cu |
The compensating filter | The electric drive (the choice of the filter), is made of Cu, 0.1mm a tolshchina - 3.0mmAl 0.2mm thickness - 3.7mmAl 0.3mm thickness - 3.8mmAl |
The size of the entering field | 150mm/100mm (6/4") / 230mm/160mm (9/6") |
Permission of the central limitation | 5.4 lp/mm (6") 5.0 lp/mm (9") 6.4 lp/mm (4") 6.0 lp/mm (6") |
System of the image | |
System of the image | 1mpks CCD camera |
Effective pixels | 1000 (IN) ×1000 (Sh) |
The allowing density | 12 bits (4,096 gradation) |
Storing of personnel | 172 images |
Processing image | Strengthening edges / Window/adjustment of level |
Monitor | |
Monitor type | 19" LCD, Color or black-and-white |
Lattice | |
Grid ratio | 10:1 |
Strips on centimeter | 40 lines/cm (Al, the standard) or 44 lines/cm (Fiber, about pts.) |
Material between space | Aluminum (standard) or Fibre (opts) |
Digital roentgenograph (option) | |
Display mode | Long / pulse / the strengthened pulse rentegoskopiya Digital spot / Serial X-ray analysis DSA in real time (option) |
Serial shooting | 0.5, 1.0, 2.0, 3.0, 3.75, 5.0, 7.5 fps 1024×1024матрица, 12 bits |
Shooting of DSA in real time (option) | 0.5, 1.0, 2.0, 3.0, 3.75, 5.0, 7.5 fps 1024×1024матрица, 12 bits |
Storage image HDD | |
Exit of media | DVD-R/CD-R |
Processing image | Strengthening of edges, Window/adjustment of level Flip (vert/horizon, black / white inversion, rotation of the image 4th zoom, Panorama Display several image 2×2/4×4 Measurement of a distance/corner Display of the summary Repeated masking (the DSA option is required.) Repeated registration (the DSA option is required.) Laying (the DSA option is required.) Processing of the RSM filter (the DSA option is required.) |