The mammography LILYUM system of the known global manufacturer of Metaltronica, Italy who is one of world leaders in production of x-ray systems for mammography today is equipped by the easy C-shaped holder and the convenient control panel, it is reliable and provides high quality of images and comfort for the patient at a low radiation dose; the additional device for a stereotaksiya with computer management.
Lilyum has the innovation system of micro processor management of exposure on means of the software of ULTRA. This standard feature does installation of mammography system incredibly fast and simple. ULTRA includes the advanced diagnostic functions, and also the loader of data for usability.
The operator will estimate brightness, and effective design of graphic LCD Lilyum displays at which all current parameters and data in a digital form are displayed:
- Angle of rotation of the With-shaped rack
- Compression force
- Breast density
- Average dose on gland
New system of a compression "μPress" which will allow to make correctly and at the same time more comfortably a compression, the safe and convenient, strengthened and steady compression depending on features of a breast with the minimum discomfort of the patient.
The maximum comfort of the patient is also reached due to modern design of a compression pedal with smooth bends and straight lines which is available in different form factors.
Technical characteristics:
POWER SUPPLY
Linear tension |
220/230/240 In alternating current of ±10% 50/60 Hz |
Linear tension (optionally) |
115 V of alternating current of ±10% 50/60 Hz |
Power |
6.6 kVA (0.5 kVA in a waiting mode) |
Absorption current |
30 A peak |
Quantity of phases |
1 or 2 it (is configured) |
Connection |
Stationary installation (IEC 60601-1 standard) |
Wall connection |
20 A a safety lock (40 A safety lock for 115 V of alternating current) or the Thermomagnetic automatic switch |
Main resistance |
<0.50 Ω |
HIGH-VOLTAGE GENERATOR
Compensation of linear tension |
Automatic The high-voltage generator with the closed cycle and the line of kV Compensation of direct feedback |
Inverter frequency |
50 kHz |
Pulsation/Amplitude frequency |
100 kHz <2% |
Power output of the generator |
5 kW (at 35 kV) |
Nominal electric power |
4.2 kW =140 ma*30 kV (3 sec.) |
Range of kV |
20 / 35 kV |
Permission of kV (Manual and automatic mode) |
0.5 kV |
DEVICE OF THE X-RAY TUBE (I.A.E. XM12)
Speed of anode rotation |
3000 rpm 50 Hz 10000 rpm 150 Hz (optionally) |
Anode mirror material |
Molybdenum |
Maximum heat content of the anode |
225 kJ (300 thousand. Those) |
Maximum speed of cooling of the anode |
750 W (60 thousand. That/min.) |
Maximum heat content of a X-ray tube |
320 kJ (425 thousand. Those) |
ISOCENTRIC WITH-SHAPED SUPPORT
Focal length |
65 cm |
Manual rotation |
±180 °C a disk brake (±135 °C a with-shaped support in the lowest situation) |
Motorized rotation (optionally) |
±180 ° (90 °/8 sec. with strengthening and linear delay) |
Rotation for projection of a biopsy |
Motorized ±15 ° microprocessor control |