Description
The system radiographic digital Blue - DR is a unique digital device where the modern digital system with the flat panel dynamic detector which considerably exceeds digital systems on the basis of AXRI (Amplifier of the X - ray Image) is used, allowing to receive high - contrast images with high resolution at a smaller exposure dose.
Advantages
- The device is intended for routine diagnostics and combines modern digital technologies and simplicity of management for a qualitative X - ray analysis
- Soft and silent work provides comfort to the patient
- Ergonomic control panel
- Fast system of identification of mistakes
- Creation and editing base of patients
- Wide choice of possible combinations. The system can be put both with ceiling, and with floor fastening of a radiator, with one detector or with two detectors, depending on your requirements and the purposes
- Parameters of a radiological survey can be adjusted according to the body weight of the patient (the Child (very small) Small Average (Usual) Large)
- Availability of the APR mode (anatomically programmable radiography)
Technical characteristics of system of radiographic digital BLUE - DR
Name |
Parameters |
X - ray generator |
|
Range of tension, kV |
50 ~ 150 kV |
Range of change of anode current |
10 ~ 630 ma |
Range of change of time of exposure |
0.001 ~ 10 sec. |
Maximum electricity consumption |
70 kVA |
Nominal parameters |
70 kV - 630ma/150kv - 250ma |
Power supply voltage |
single - phase, Alternating current of 220 V ± 10% , 50/60 Hz |
X - ray tube |
|
Maximum tension of a tube |
150 kV |
Own filtering |
At least an equivalent 0.7 mmal at 75 kV |
Type |
The rotating anode |
Amount of focus |
small focus: 0.6 mm / big focus: 1.2 mm |
Heat capacity of the anode |
300 KTE |
Physical characteristics X - ray tube |
|
Dimensions
|
Total length: 476 mm Maximum diameter: 152.4 mm |
Target
|
Corner: 12 degrees tube material: rhenium - tungsten |
Protection against radiation (According to requirements of IEC 60601 - 1 - 3) |
Radiation leak indicator: 150 kV 3.4 ma
|
Focus of the image
|
354 × 354 mm at distance a source image of 970 mm |
Weight (pribliz. ) |
18 kg |
Plugs of a high voltage |
According to requirements of IEC 60526 |
Way of cooling |
Natural or forced air |
Absolute maximum and minimum nominal parameters. (Exceeding of these sizes, is not allowed under no circumstances) |
|
Maximum tension of a tube |
150 kV |
Maximum tension |
78 kV |
Minimum tension of a tube |
40 kV |
Maximum current of a tube
|
Big focus: 1000 ma Small focus: 400 ma |
Maximum current of heat
|
Big focus: 5.5 A Small focus: 5.2 A |
Heat tension
|
Large focus (at max. current of heat 5.5 A): from 12.7 to 17.1 In Small focus (at max. current of heat 5.2 A): from 7.0 to 9.4 |
Frequency range |
from 0 to 25 kHz |
Average input power: 142 W (200 THOSE / c) |
radiographic |
Thermal characteristics
|
Heat capacity of the anode: 210 kJ (300 KTE) Maximum anode intensity of heatdispersion: 475 W (667 THOSE / c) Heat capacity of the body: 900 kJ (1250 KTE) |
Maximum intensity of heatdispersion of the body |
Without air circulation pulser: 180 W (15 KTE/min. ) |
Environment conditions |
|
Operating conditions |
Temperature: from 10 to 40 ºC Humidity: from 30 to 85% Atmospheric pressure: from 70 to 106 kPa |
Conditions of transportation and storage
|
Temperature: from - 20 to 70 ºC Humidity: from 20 to 90% Atmospheric pressure: from 50 to 106 kPa |
Collimator |
|
Filtering |
1.5 mmal |
Maximum tension |
150kvp (peak anode tension) |
Maximum size of the field |
35 X 35 cm or more (Distance source image of 65 cm) |
Minimum size of the field |
5 X 5 cm or less (Distance source image of 100 cm) |
Illumination (100 cm) |
160 Luxury and more |
Lamp operating time |
30 sec. ± 5% |
Detector |
|
Detector type |
The detector with the flat panel |
Area of diagnostics |
425х425 mm |
Pixel step |
168 microns |
Permission |
3268х3268 |
Analog - digital converter |
14 bits |
Data processing speed |
Less than 4.5 sec. after a radiation X - ray |
Complex configuration |
Control unit (Alternating current 100 ~ 220±10% , 50/60 Hz), Distribution box |
Processing of images. Requirements to the personal computer |
|
Operating system |
Window XP |
Processor |
GHz Pentium 4, 2.0 (at least) |
Random access memory |
1 Gigabyte (at least) |
Hard drive |
60 Gigabyte (at least) |
Monitor
|
the color monitor with the resolution of 32 bit (the liquid crystal monitor 18.1 inches or the electron beam monitor of 19 inches) |
CD/R |
48 of X (at least) |
Permission |
1280 x 1024 |
Network |
the card for the high - speed Ethernet network |
Fastening of a X - ray tube |
|
Support of a X - ray tube ceiling |
|
Vertical movement |
1350 mm ± 10% |
Longitudinal movement |
(2700 ~ 4000) mm ± 10% |
Lateral movement |
(1400 ~ 2000) mm ± 10% |
Rotation of a rack |
± 180 ° |
Rotation of a X - ray tube |
±90 ° |
Power supply |
Alternating current of 220 V ± 10% , 50/60 Hz |
Support of a X - ray tube floor (vertical) |
|
Vertical movement |
1300 mm ± 10% |
Longitudinal movement |
(2100 ~ 3000) mm ± 10% |
Lateral movement |
350 mm ± 10% |
Rotation of the body |
±90 ° |
Rotation of a X - ray tube |
±180 ° |
Entrance tension |
Direct current of 24 V ± 10% |
Digital rack of pictures |
|
The rack is vertical detector multi position |
|
Vertical movement |
1165 mm ± 10% |
Increase in speed |
32 mm / with ± 10% |
Reduction of speed |
34 mm / with ± 10% |
Detector tilt angle |
+ 90 ° ~ - 15 ° |
Speed of turn of the detector |
34 mm / with ± 10% |
Power supply |
Alternating current of 220 V ± 10% , 50/60 Hz |
Detector support |
|
Vertical movement |
1100 mm ± 10% |
Entrance tension |
Alternating current 16 ~ 26B |
Table |
|
Table of the detector 6 directions (option) |
|
Material of a working surface |
Acryle or Melamine |
Dimensions of a working surface |
(2000 x 750) mm ± 10% |
Vertical movement |
290 mm ± 10% |
Increase in speed |
13 mm / with ± 10% |
Reduction of speed |
15 mm / with ± 10% |
Longitudinal movement |
270 mm ± 10% |
Lateral movement |
120 mm ± 10% |
Promotion of the cartridge for the detector |
260 mm ± 10% |
Power supply |
Alternating current of 220 V ± 10% , 50/60 Hz |
Table of the detector 4 directions |
|
Material of a working surface |
Acryle or Melamine |
Dimensions of a working surface |
(2000 x 750) mm ± 10% |
Longitudinal movement |
270 mm ± 10% |
Lateral movement |
120 mm ± 10% |
Promotion of the cartridge for the detector |
260 mm ± 10% |
Power supply |
Alternating current of 110 V ± 10% , 50/60 Hz |
Table mobile |
|
Surface size |
(2000 x 735) mm ± 10% |